Horiba OCMA-550 # 3200577566 Oil Content Monitor / Analyzer, Old # 30005433
The OCMA-550 measures residual oil on components and concentrations of oil adhered on solids such as soil
Measurement can be easily made only by injecting the extracted water sample into the attached cell and setting it to the equipment
The Horiba OCMA-550 # 3200577566 Oil Content Monitor / Analyzer, Old # 30005433 measures residual oil on components and concentrations of oil adhered on solids such as soil. Measurement can be easily made only by injecting the extracted water sample into the attached cell and setting it to the equipment. The Horiba OCMA-550 # 3200577566 Oil Content Monitor features a simple design which allows opening/closing of the door to setting of cell with just one hand. This is best for measurement of extracted samples such as evaluation of residual oil on components and measurement of oil contained in food.

Kit Contents
- Horiba OCMA-550 # 3200577566 Oil Content Monitor / Analyzer
- Standard Accessories:
- Dropper Made of Polyethylene, 2.5 mL
- Code set Power Supply Cable
- B-Heavy Oil 10 mL
- Cell Quartz (20 mm)
- Cell Cap
- Instruction Manual
* Horiba S-316 Extraction Solvent Oil not Included
For the First Purchase Customer
In order to measure oil content with Horiba OCMA-550 Oil Content Monitor / Analyzer, you need the following products. If you don't have these products, please purchase from optional list.
Must items for measurement
Recommended options (Not including maintenance parts)
*Oils containing isoalkanes may not be removed by Solvent Reclaimer.
Features
Quick and easy measurement in approximately 3 minutes*1 Measurements can be made easily and quickly by simply pressing the start button. Compared to the n-hexane method, measurement times can be significantly reduced. *Excluding time for activities such as warming, calibration, rinsing | Capable of measuring low-boiling point oil components The n-hexane method requires the evaporation of the oil extraction solvent, which leads to volatilization of low boiling point oils (such as gasoline). The OCMA-500 series eliminates the need for an evaporation process, allowing for the prevention of volatilization in low-boiling point oil components | Oil content can be displayed as a concentration (value) Unlike the n-hexane method, which calculates concentration by measuring the weight of oil content, OCMA-550 can directly display oil concentration. *Please contact us for inquiries regarding the types of oil that can be detected. | Environmental-friendly, safe, and cost-effective oil extraction solvents are in use The domestic model OCMA-505-H/555-H uses solvent H-519 for oil extraction, while the international model OCMA-500/550 uses solvent S-316. Both H-519 and S-316 are not regulated under the Montreal Protocol (Ozone Layer Protection Act) and the PRTR system (Act on the Evaluation of Chemical Substances and Regulation of Their Manufacture, etc. (CSCL)). Furthermore, compared to the traditional solvent H-997, solvent H-519 has achieved similar regeneration efficiency to solvent S-316, allowing for a reduction in running costs. |
Specifications
| Model | OCMA-550 |
| Product name | Oil content analyzer |
| Measurement method | Solvent extraction - non-dispersive infrared absorption analysis method |
| Measured objects | Substances extracted from sample water into solvent and having infrared absorption near a wavelength from 3.4 μm to 3.5 μm |
| Measurement range | 0 mg/L to 200 mg/L |
| Resolution | For mg/L 0 to 99.9: 0.1, 100 to 200: 1 For mg/g, mg/kg, mg/PC 0 to 9.99: 0.01, 10.0 to 99.9: 0.1, 100 to 200: 1 Abs.: 0 to 1.000 Abs.: 0.001 Abs. |
| Repeatability | 0 mg/L to 9.9 mg/L: ±0.4 mg/L ±1 dig. 10.0 mg/L to 99.9 mg/L: ±2.0 mg/L ±1 dig. 100 mg/L to 200 mg/L: ±4 mg/L ±1 dig. * For standard liquids |
| Display method | 3.5 inches, 320×240 dots Backlight Color graphic LCD |
| Calibration method | Select each optionally zero calibration and span calibration |
| Extraction solvent | S-316 |
| Amount of extraction solvent required | Approx. 6.5 mL |
| Extraction method | Using the extraction solvent, and extracted manually outside the product |
| Ambient operating temperature | 0°C to 40°C (no condensation) |
| Power supply | 100 V to 240 V AC ±10%, 50/60 Hz |
| Power consumption | 100 V to 240 V AC, Approx. 60 VA |
| External dimensions | 195 (H) × 253 (W) × 293 (D) mm |
| Mass | Approx. 5 kg |
| External output | Output to a USB memory stick |
| Functions | 300-item data memory (measurement history) Stabilized measurement value display Self error determination Clock Conversion of measurement units Compliant with ASTM D7066-4 |
Applications
Residual oil components - For quality control of components
| For piping of air conditioners and refrigerator - To prevent reduction of cooling function
| Semiconductor field - For evaluation of degreasing capability
| Others - Useful for soil (environmental pollution), food(health hazard) and gas(quality deterioration)
|
Related Products
Horiba S-316 Extraction Solvent Oil - 5200100690
This solvent is required to:
- Prepare the calibration solution to calibrate the infrared Spectrometer OCMA: pure solvent (zero Calibration), solvent + oil (span calibration)
- Extract the oil and grease from the sample
Downloads