Measuring range: 0.8 to 80,000 mg/L Br- | (10-5 to 1 mol/L) Br-
pH Range: 800 mg/L (10-2 mol/L) Br- | pH 1.5 to 11.5
Applicable temperature range: 0 to 50℃

Bromide ion electrode 8005-10C.


line
Specifications
line

Measuring range

0.8 to 80,000 mg/L Br-
(10-5 to 1 mol/L) Br-

pH range

800 mg/L (10-2 mol/L) Br-
pH 1.5 to 11.5

Applicable temperature range

0 to 50℃

Response time (90%)

Within 5 seconds

Applicable reference electrode

2565A

Selection coefficient

S2O32-、l-、S2-、CN- = Not acceptable
MnO4- = 1
Cl-、PO42- = 100
F-、NO3-、SO42- = 1,000

The response time is the time which is required to reach 90% response when the ion concentration is gradually changed from 10-4 mol/L to 10-2 mol/L with the solution stirred.
Exception:
*1: 90% response when ion concentration is changed to 10-6 mol/L ∼ 10-2 mol/L
*2: 95% response when ion concentration is changed to 10-3 mol/L ∼ 10-1 mol/L
*3: 95% response when ion concentration is changed to 10-4 mol/L ∼ 10-2 mol/L
*4: 95% response when ion concentration is changed to 10-4 mol/L ∼ 10-1 mol/L

• The selection coefficient is a ratio of the limit concentration of coexisting ions (mol/L) to the ion concentration to be measured (mol/L); The value of 1000 means that the coexisting ions can be permitted up to 1000 times the ion measured and ''not acceptable'' means that chemical change occurs in the solid response membrane.